6
MEASure Subsystem
MEASure:DIODe?
150
U3606A Programmer’s Reference
MEASure:DIODe?
Syntax
MEASure:DIODe?
This command first resets all diode test parameters and trigger
parameters to their default values. Then, it configures the instrument for
diode tests and immediately triggers the test. A single reading is returned.
Remarks
•
The range and resolution are fixed for diode tests. The range is 1 VDC
(with a 1 mA current source output) and the resolution is set to
4½ digits.
•
The voltage is displayed on the front panel if it is in the 0 V to 1.2 V
range. The meter beeps when the signal transitions to the 0.3 V to
0.8 V threshold (unless beep is disabled). If the signal is greater than
1.2 V, “OPEn” is displayed on the front panel.
•
The
MEASure:DIODe?
command returns the measured voltage,
regardless of its value.
Example
&
MEAS:DIOD?
This command configures the instrument for
continuity measurements, triggers the
instrument to take a reading, and then transfers
the reading to the instrument output buffer.
$
+1.321300E-01
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Страница 2: ......
Страница 3: ......
Страница 13: ...U3606A Programmer s Reference XI Device specific errors 341 Self test errors 342 Calibration errors 343 ...
Страница 14: ...XII U3606A Programmer s Reference ...
Страница 27: ...Introduction to SCPI 1 SCPI Status System U3606A Programmer s Reference 13 Figure 1 1 Status system diagram ...
Страница 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Страница 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Страница 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Страница 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Страница 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Страница 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Страница 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Страница 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Страница 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Страница 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Страница 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...