CONFigure Subsystem
4
CONFigure:FREQuency
U3606A Programmer’s Reference
113
•
If the input signal is greater than can be measured on the selected
range (manual ranging), the instrument gives an overload indication.
•
Positive overload: “
+9.37
” or “OL” from the front panel.
•
Negative overload: “
-9.37
” or “–OL” from the front panel.
•
If a range change is in progress (due to the autoranging setting) when
you query the instrument, the value “
+9.37
” will be
returned. Wait for the instrument to select an appropriate range before
querying the instrument again.
Example
See also
“CONFigure[:VOLTage]:AC” on page 84
“CONFigure:CURRent:AC” on page 93
“INITiate[:IMMediate]” on page 120
&
CONF:CURR:AC
This command configures the instrument for
AC current measurements. The default range
(autorange) and resolution (5½ digits) are used.
&
CONF:FREQ
This command configures the instrument for
frequency measurements via the current path.
The range and resolution follows the
configuration of the AC current function.
&
READ?
This command places the instrument in the
“wait-for-trigger” state, triggers a
measurement, and sends the reading to the
instrument memory and output buffer.
$
+1.02
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Страница 2: ......
Страница 3: ......
Страница 13: ...U3606A Programmer s Reference XI Device specific errors 341 Self test errors 342 Calibration errors 343 ...
Страница 14: ...XII U3606A Programmer s Reference ...
Страница 27: ...Introduction to SCPI 1 SCPI Status System U3606A Programmer s Reference 13 Figure 1 1 Status system diagram ...
Страница 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Страница 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Страница 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Страница 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Страница 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Страница 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Страница 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Страница 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Страница 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Страница 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Страница 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...