1
Calibration Procedures
24
Keysight U1270 Series Service Guide
Calibration Process
1
Prior to performing the verification tests, see the
2
Perform the verification tests to characterize the instrument; see
3
Unsecure the instrument for calibration; see
4
Prior to performing the adjustments, see the
5
Perform the adjustment procedure; see
6
Secure the instrument against unauthorized calibration; see
on page 46. Ensure that the instrument has quit the
adjustment mode and is turned off.
7
Record the new security code and calibration count in the instrument's
maintenance records.