Calibration Procedures
1
Keysight U1270 Series Service Guide
43
ACmA/ACA
30 mA
03.000 mA (70 Hz)
0.9 to 1.1
×
Reference value
30.000 mA (70 Hz)
0.9 to 1.1
×
Reference value
300 mA
030.00 mA (70 Hz)
0.9 to 1.1
×
Reference value
300.00 mA (70 Hz)
0.9 to 1.1
×
Reference value
3 A
0.3000 A (70 Hz)
0.9 to 1.1
×
Reference value
3.0000 A (70 Hz)
0.9 to 1.1
×
Reference value
10 A
3.0000 A (70 Hz)
0.9 to 1.1
×
Reference value
10.000 A (70 Hz)
0.9 to 1.1
×
Reference value
Capacitance
OPEN
OPEN
With cables connected to CAP/COM terminals
10 nF
04.000 nF
0.9 to 1.1
×
Reference value
10.000 nF
0.9 to 1.1
×
Reference value
100 nF
010.00 nF
0.9 to 1.1
×
Reference value
100.00 nF
0.9 to 1.1
×
Reference value
1000 nF
0100.0 nF
0.9 to 1.1
×
Reference value
1000.0 nF
0.9 to 1.1
×
Reference value
10
μ
F
10.000
μ
F
0.9 to 1.1
×
Reference value
100
μ
F
100.00
μ
F
0.9 to 1.1
×
Reference value
1000
μ
F
1000.0
μ
F
0.9 to 1.1
×
Reference value
10 mF
10.000 mF
0.9 to 1.1
×
Reference value
OPEN
OPEN
OPEN terminals (no cables connected to CAP/
COM terminals)
[a]
Table 1-3
Adjustment input values (continued)
Test function
Step
Reference value
Valid reference input
Содержание U1271A
Страница 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Страница 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 53: ...Service and Maintenance 2 Keysight U1270 Series Service Guide 53 Figure 2 1 Testing Fuse 1 Smart Smart ...
Страница 54: ...2 Service and Maintenance 54 Keysight U1270 Series Service Guide Figure 2 2 Testing Fuse 2 Smart Smart ...
Страница 56: ...2 Service and Maintenance 56 Keysight U1270 Series Service Guide Figure 2 3 Replacing Fuse 1 1 2 3 ...
Страница 57: ...Service and Maintenance 2 Keysight U1270 Series Service Guide 57 Figure 2 4 Replacing Fuse 2 1 2 3 ...