Calibration Procedures
46
Keysight U1270 Series Service Guide
7
Use the arrow keys to enter the actual applied input values.
8
Press
to start the adjustment.
CAL
is shown on the display to indicate
that the calibration is in progress.
9
Upon completion of each adjustment value, the display will show
PASS
. If the
adjustment fails, the instrument will sound a long beep and the calibration
error number is shown on the display.
10
Turn the rotary switch to the next function according to the
Test function
column shown in
. Repeat
for each adjustment point
shown in the adjustment table.
11
Verify the adjustments using the
“Performance Verification Tests”
Exiting the adjustment mode
1
Remove all the shorting plugs and connectors from the instrument.
2
Record the new Calibration Count.
3
Press
and
simultaneously to exit the Adjustment Mode.
4
Cycle the instrument’s power. The instrument will then be secured.
NOTE
You are highly recommended to complete the adjustments in the same order as
shown in
.
Hz % ms
Log
NOTE
If the adjustment fails, check the input value, range, function, and entered
adjustment value before repeating the adjustment steps.
Shift
View
Esc
Auto
Trig
Hold
Содержание U1271A
Страница 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Страница 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 53: ...Service and Maintenance 2 Keysight U1270 Series Service Guide 53 Figure 2 1 Testing Fuse 1 Smart Smart ...
Страница 54: ...2 Service and Maintenance 54 Keysight U1270 Series Service Guide Figure 2 2 Testing Fuse 2 Smart Smart ...
Страница 56: ...2 Service and Maintenance 56 Keysight U1270 Series Service Guide Figure 2 3 Replacing Fuse 1 1 2 3 ...
Страница 57: ...Service and Maintenance 2 Keysight U1270 Series Service Guide 57 Figure 2 4 Replacing Fuse 2 1 2 3 ...