Calibration Procedures
1
Keysight U1270 Series Service Guide
41
Valid adjustment input values
Adjustment can be accomplished using the following input values below.
Table 1-3
Adjustment input values
Test function
Step
Reference value
Valid reference input
DCmV
SHORT
SHORT
SHORT V/COM terminals
30 mV
30.000 mV
0.9 to 1.1
×
Reference value
300 mV
300.00 mV
0.9 to 1.1
×
Reference value
ACmV
30 mV
3.000 mV (70 Hz)
0.9 to 1.1
×
Reference value
30.000 mV (70 Hz)
0.9 to 1.1
×
Reference value
30.000 mV (30 kHz)
0.9 to 1.1
×
Reference value
300 mV
30.00 mV (70 Hz)
0.9 to 1.1
×
Reference value
300.00 mV (70 Hz)
0.9 to 1.1
×
Reference value
300.00 mV (30 kHz)
0.9 to 1.1
×
Reference value
DCV
SHORT
SHORT
SHORT V/COM terminals
3 V
3.0000 V
0.9 to 1.1
×
Reference value
30 V
30.000 V
0.9 to 1.1
×
Reference value
300 V
300.00 V
0.9 to 1.1
×
Reference value
1000 V
1000.0 V
0.9 to 1.1
×
Reference value
Содержание U1271A
Страница 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Страница 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 53: ...Service and Maintenance 2 Keysight U1270 Series Service Guide 53 Figure 2 1 Testing Fuse 1 Smart Smart ...
Страница 54: ...2 Service and Maintenance 54 Keysight U1270 Series Service Guide Figure 2 2 Testing Fuse 2 Smart Smart ...
Страница 56: ...2 Service and Maintenance 56 Keysight U1270 Series Service Guide Figure 2 3 Replacing Fuse 1 1 2 3 ...
Страница 57: ...Service and Maintenance 2 Keysight U1270 Series Service Guide 57 Figure 2 4 Replacing Fuse 2 1 2 3 ...