5
Performance Tests and Calibration
68
Keysight U1241AN User’s and Service Guide
Calibration Security
The calibration security code prevents accidental or unauthorized adjustments to
the multimeter. The multimeter is secured when the multimeter is shipped from
factory. Before performing any adjustment to the multimeter, you are required to
unsecure the multimeter by entering the correct security code (see
the Multimeter for Calibration”
on page 68). The security code may contain up to
4 numeric characters.
Unsecuring the Multimeter for Calibration
Before performing adjustment to the multimeter, you are required to unsecure the
multimeter by entering the correct security code. The security code is set to 1234
when the multimeter is shipped from the factory. The security code is stored in
non-volatile memory, and does not change when power is off.
Unsecuring the multimeter from the front panel
1
Turn the rotary switch to
.
2
Press
and
simultaneously to enter the Calibration Security Code
entry mode. The primary display shows
5555
and the secondary display
shows
SEU
.
3
Press
or
to step each character in the code. Press
or
to change
the value of the selected character.
4
Press
(Save) when done.
5
If the correct security code is entered, the secondary display will show “
PAS
”.
If an invalid code is entered, the multimeter will shows error code “
E02
” on the
secondary display for approximately 3 seconds before returning to the
Calibration Security entry mode.
NOTE
You can unsecure and change the security code from the multimeter front panel.
If you forgot your security code, see
Содержание U1241AN
Страница 1: ...Keysight U1241AN Handheld Digital Multimeter User s and Service Guide ...
Страница 14: ...14 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 28: ...1 Getting Started 28 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...