5
Performance Tests and Calibration
66
Keysight U1241AN User’s and Service Guide
5
Turn the rotary switch to
position
100
Ω
100.00
Ω
± 0.33
Ω
[a]
1000
Ω
1000.0
Ω
± 3.3
Ω
10 k
Ω
10.000 k
Ω
± 33
100 k
Ω
100.00 k
Ω
± 330
Ω
1000 k
Ω
1000.0 k
Ω
± 3.3 k
Ω
10 M
Ω
10.000 M
Ω
± 62 k
Ω
100 M
Ω
100.00 M
Ω
± 1.53 M
Ω
[b]
6
Turn the rotary switch to
position
1000 nF
1000.0 nF
± 10.3 nF
10 µF
10.000 µF
± 0.103 µF
100 µF
100.00 µF
± 1.03 µF
1000 µF
1000.0 µF
± 10.3 µF
10 mF
10.000 mF
0.103 mF
7
Turn the rotary switch to
position
1000 µA
1000.0 µA
± 2.3 µA
10000 µA
10000 µA
± 23 µA
8
Press
to go to
function
1000 µA
1000.0 µA, 2 kHz
± 10.2 µA
10000 µA
10000 µA, 2 kHz
± 102 µA
9
Turn the rotary switch to
position
100 mA
100.00 mA
± 0.23 mA
440 mA
400.0 mA
[c]
± 1.1 mA
10
Press
to go to
function
100 mA
100.00 mA, 2 kHz
± 1.02 mA
440 mA
400 mA
, 2 kHz
± 4.2 mA
11
Turn the rotary switch to
position
10 A
10.000 A
[d]
± 33 mA
12
Press
to go to
function
10 A
10.000 A
± 102 mA
13
Turn the rotary switch to
position
100 mV
100.00 mV
± 0.11 mV
Table 5-2
Verification tests (continued)
Step
Test Function
Range
5520A Output
Error from nominal
1 year
U1241AN
Содержание U1241AN
Страница 1: ...Keysight U1241AN Handheld Digital Multimeter User s and Service Guide ...
Страница 14: ...14 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 28: ...1 Getting Started 28 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...