Performance Tests and Calibration
5
Keysight U1241AN User’s and Service Guide
61
Input-A Terminal Test
This test determines if the input warning of the
A
current terminal test is
functioning properly. The multimeter sounds an alerting beep when the test lead
is inserted to the
A
input terminal but the rotary switch is not set to the
corresponding
A
location. The display indicates a flashing
AErr
annunciator until
the test lead is removed from the
A
input terminal.
Input-mA Terminal Alert Test
This test determines if the detected input voltage level is within the accepted
boundaries when the
μ
A/mA
input terminal is connected. The multimeter sounds
an alerting beep when the
μ
A/mA
input terminal detects a voltage level of higher
than 1.6 V. The display indicates a flashing
CErr
annunciator until the test lead is
removed from the
μ
A/mA
input terminal.
NOTE
The alerting beep will continue to function even if the beep function is disabled.