Installation Note N5242-90010 31
Description of the Upgrade
Instruments with serial prefixes <5200 only
Instruments with serial prefixes <5200 only
Step 14. Modify the N5245-60124 Noise Figure Board (With Tabs) Per
Instructions in the Upgrade Kit N5245-60125 (Serial prefixes <5200
Only)
Follow the instructions in your N5245-60125 kit to modify your N5245-60124
to a no tabs noise board. Refer to
http://literature.cdn.keysight.com/litweb/pdf/N5245-90125.pdf
.
All Instruments
Step 15. Install the A7 Noise Receiver Board
Refer to
for this part of this step of the procedure. Although only
Option 419 is shown in the illustration, Option 423 is similar in appearance.
New parts are listed in
.
1. Place the analyzer top-side up on a flat surface.
2. Insert the A7 noise receiver board in the analyzer as shown. Make sure it is
fully seated in the motherboard connector.
If your PNA serial prefix is >5200, skip to the next step
.