16 Installation Note N5242-90010
Description of the Upgrade
Windows 7 Operating System (Windows 7 OS) Only
-
Windows 7 Operating System (Windows 7 OS) Only
Step 4. Update Firmware with A.10.47 or above
1.
If the PNA is running, click
File
>
Exit
2.
http://na.support.keysight.com/pna/firmware/firmware.htm
and
download the latest PNA firmware.
3.
Click on the firmware to be downloaded.
4.
Save the program to disk (hard drive of your PC or to your PNA).
5.
If necessary, transfer the file from your PC to your PNA using LAN, CD, or
USB Pen drive.
6.
Double-click the file on the PNA. Follow the prompts to complete the
firmware installation.
7.
If the PNA application does not start up automatically, double-click the
network analyzer icon on the desktop.
Step 5. Verify Windows 7 OS with New A.10.47.00 Firmware or
Greater
1.
After the new firmware is installed, verify the firmware operates without
errors.
2.
After the instrument powers up, on the toolbar click or press
Help > About
NA
to verify the firmware has been updated to A.10.47.00 or greater.
3.
Power down your instrument.
If your PNA is already running firmware A.10.47 or above, skip this step to
“Step 6. Remove the Front Panel Assembly” on page 17
.
If you have problems with updating the PNA’s firmware, refer to
Assistance from Keysight” on page 4
If you have already verified that your PNA is running firmware A.10.47 or
above, skip this step to
“Step 6. Remove the Front Panel Assembly” on
.
If you have problems with updating the PNA’s firmware, refer to