2
Safety and Regulatory Information
20
N7040A/1A/2A Rogowski AC Current Probes User Guide
Pre-Use Safety Checks and Warnings
WARNING
Indoor Use Only. Do not operate in wet / damp environments. Keep product
surfaces dry and clean.
Do not operate in an explosive environment.
CAUTION
The intended use of these probes is for the measurement of AC current.
Keysight Technologies accepts no responsibility for any damage resulting
from careless use, or failure to observe these instructions.
WARNING
If HAZARDOUS LIVE voltages are present and accessible in the installation,
appropriate protective measures must be taken to avoid electric shock.
WARNING
The integrity of the insulation around the Rogowski coil must be VISUALLY
INSPECTED before and after each use, and the probe must NOT BE USED if
there are signs of coil damage.
WARNING
Be extremely careful when using the coil of the probe on uninsulated
conductors. The power of the uninsulated conductor must be turned off
when applying or removing the probe’s coil from this conductor.
In addition, make sure that the probe is never used at voltages greater than
its voltage rating (safe PEAK working voltage) which is clearly labeled on
the probe cable adjacent to the coil. For N7040A and N7041A, it is 5kV peak
and for N7042A, it is 1.2kV peak.
WARNING
Voltage ratings are only valid if the ‘free end’ of the coil is fully inserted into
the probe’s ferule, and remains fully inserted during use. Refer to the topic
“Probe Handling Precautions"
on page 16 to know more.
Содержание N7040A
Страница 1: ...Keysight N7040A 1A 2A Rogowski AC Current Probes User Guide ...
Страница 18: ...1 Overview 18 N7040A 1A 2A Rogowski AC Current Probes User Guide ...
Страница 40: ...5 Returning the N7004A for Repair or Service 40 N7040A 1A 2A Rogowski AC Current Probes User Guide ...
Страница 41: ...N7040A 1A 2A Rogowski AC Current Probes User Guide 41 ...