Keysight N5990A Type-C Multi-Standard Solution
User Guide
7
Appendix
For Standard combination USB-RX & USB-TX
For Standard combination USB-RX & USB-TX & DP-TX
The USB-TX tests with the Keysight Type-C fixture N7015A require the use
of transfer functions.
To get comparable USB RX-test results with the USB-IF fixture (left) and
the Keysight N7015A fixture (right), USB reference channel must be
embed the into the measurement signal and de-embed the N7018A
fixture.
Figure 26