Keysight N5990A Type-C Multi-Standard Solution User Guide
5
Calibrations Procedures
Figure 18
Transfer Function File
Transfer function generation
N5990A comes with the examples of transfer functions, which are
measured and calculated for one specific setup. You can copy those
transfer functions from your ValiFrame PC to the scope:
Figure 19
Default Transfer Functions
But you must be aware that your own setup may have different properties.
For proper tests, you should measure S-parameters of your own setup and
calculate the transfer functions. In the appendix, you can find the tables of
the used S-parameter files for the USB reference channel and the Keysight
fixture N7015A.