
34
N5245-90002 User’s and Service Guide
N5244/45AS Option H85
Example: Making High Power Measurements With Option H85
Final Setup
21.Verify that all of the power and attenuator settings are correct.
22.In the
Response
drop-down menu select
Measure
>
S21
.
Figure 31
Response Menu
23.To Perform a response calibration following steps a, b, c and d:
a. Connect the test port cables of the analyzer to form a thru configuration.
b. In the
Response
drop-down menu select
Cal Wizard
(see
). In the dialog box,
select
UNGUIDED Calibration (Response, 1-port 2-port): Use Mechanical Standards
>
Next.
Figure 32
Calibration Wizard
Содержание N5244/45AS Option H85
Страница 4: ...Contents ii Contents ...
Страница 5: ...N5245 90002 User s and Service Guide 1 N5244 45AS Option H85 ...
Страница 13: ...N5245 90002 User s and Service Guide 9 N5244 45AS Option H85 Operation Figure 9 Standard 4 Port Front Panel ...
Страница 50: ...46 N5245 90002 User s and Service Guide N5244 45AS Option H85 Keysight Support Services and Assistance ...