30
N5245-90002 User’s and Service Guide
N5244/45AS Option H85
Example: Making High Power Measurements With Option H85
Selecting Power Ranges and Attenuator Settings
12.Select a power range that will not exceed the maximum estimated power level, but will force
the DUT into compression. For example; if your booster amplifier has a gain of +15 dB, the
DUT will compress if supplied with +15 dBm. You may need to adjust the analyzer output
power to not exceed 0 dBm.
13.In the
Stimulus
drop-down menu select
Power
>
Power and Attenuators.
In the dialog box,
clear the
Auto Range
box
and type
[10 dB]
into
Atten. Control
for
Port 1.
Verify that
Port
Powers Coupled
is checked to ensure that Ports 1 and Port 2 power levels are the same.
Press
OK
.
Figure 26
Attenuator Control