Installation Note N5225-90107 31
Description of the Upgrade
Installation Procedure for the Upgrade
"B" Model Option Verification Procedure
1.
Start the Network Analyzer program.
2.
Once the Network Analyzer program is running:
—
Press
Help
>
About NA
and verify that Option 419 or 423 is listed in
the PNA application.
3.
After successful installation of all upgrades, some features require some
adjustments to ensure the instrument meets its specified performance.
Refer to the following Web site:
http://na.support.keysight.com/pna
Step 19.Perform Post-Upgrade Adjustments and Calibration
Adjustments
The following adjustments must be made due to the hardware changes of the
analyzer.
— source adjustment
— IF gain adjustment
— receiver characterization
— receiver adjustment
— IF Response adjustment (For A model: Options 090, 093, or 094 Only. For B
models: Options S93090xA/B, S93093A/B, or S93094A/B Only.)
— Noise adjustment (For N5244/5B models: Option 029 with S93029A/B
Only.)
These adjustments are described in the PNA Service Guide and in the PNA
on-line HELP. A list of equipment required to perform these adjustments is also
found in the service guide.
To view this service guide information, click the Chapter 3 bookmark “Tests
and Adjustments” in the PDF Service Guide
1
.
After the specified adjustments have been performed, the analyzer should
operate and phase lock over its entire frequency range.
If if the option(s) have not been enabled or if Option 401 has not been
removed, contact Keysight Technologies. Refer to
1. See