M9380A Troubleshooting
Status LED States
Once you open the M9380A SFP, select all relevant modules, and then choose Connect. The modules are ini-
tialized. Until a module is initialized, the LEDs are off.
Module
Green
Orange
Red
Off
M9300A
The Soft Front Panel has
initialized the connection
to the module
n/a
Indicates that
the VCXO is
unlocked.
l
Not connected by the SFP.
l
Failure in the power supplies. Module hard-
ware health can't be determined until the
power supply failure is resolved.
M9301A
The Soft Front Panel has
initialized the connection
to the module
Tuning is in progress, or the
M9301A is unlocked from
the reference.
n/a
l
Not connected by the SFP.
l
Failure in the power supplies. Module hard-
ware health can't be determined until the
power supply failure is resolved.
M9310A
The Soft Front Panel has
initialized the connection
to the module
n/a
n/a
l
Not connected by the SFP.
l
Failure in the power supplies. Module hard-
ware health can't be determined until the
power supply failure is resolved.
M9380A Soft Front Panel Self Tests
Start the SFP self tests by selecting: Utilities > Self Test > Run Self Test. Once completed, a dialog box
appears and reports on each of the modules, indicating if it passed or failed.