Safety Information

The following general safety precau-

tions must be observed during all 

phases of operation of this instrument. 

Failure to comply with these precau-

tions or with specific warnings or oper-

ating instructions in the product 

manuals violates safety standards of 

design, manufacture, and intended use 

of the instrument. Keysight Technolo-

gies assumes no liability for the cus-

tomer's failure to comply with these 

requirements. 

General

Do not use this product in any manner not 

specified by the manufacturer. The protec-

tive features of this product must not be 

impaired if it is used in a manner specified in 

the operation instructions.

Before Applying Power

Verify that all safety precautions are taken. 

Make all connections to the unit before 

applying power. Note the external markings 

described under “Safety Symbols”.

Ground the Instrument

Keysight chassis’ are provided with a 

grounding-type power plug. The 

instrument chassis and cover must be 

connected to an electrical ground to 

minimize shock hazard. The ground pin 

must be firmly connected to an electri-

cal ground (safety ground) terminal at 

the power outlet. Any interruption of 

the protective (grounding) conductor 

or disconnection of the protective 

earth terminal will cause a potential 

shock hazard that could result in per-

sonal injury.

Do Not Operate in an Explosive 

Atmosphere

Do not operate the module/chassis in 

the presence of flammable gases or 

fumes.

Do Not Operate Near Flammable 

Liquids

Do not operate the module/chassis in 

the presence of flammable liquids or 

near containers of such liquids.

Cleaning

Clean the outside of the Keysight mod-

ule/chassis with a soft, lint-free, 

slightly dampened cloth. Do not use 

detergent or chemical solvents.

Do Not Remove Instrument Cover

Only qualified, service-trained person-

nel who are aware of the hazards 

involved should remove instrument 

covers. Always disconnect the power 

cable and any external circuits before 

removing the instrument cover.

Keep away from live circuits

Operating personnel must not remove 

equipment covers or shields. Proce-

dures involving the removal of covers 

and shields are for use by service-

trained personnel only. Under certain 

conditions, dangerous voltages may 

exist even with the equipment 

switched off. To avoid dangerous elec-

trical shock, DO NOT perform proce-

dures involving cover or shield removal 

unless you are qualified to do so.

DO NOT operate damaged 

equipment

Whenever it is possible that the safety 

protection features built into this prod-

uct have been impaired, either through 

physical damage, excessive moisture, 

or any other reason, REMOVE POWER 

and do not use the product until safe 

operation can be verified by service-

trained personnel. If necessary, return 

the product to an Keysight Technolo-

gies Sales and Service Office for ser-

vice and repair to ensure the safety 

features are maintained.

DO NOT block the primary 

disconnect

The primary disconnect device is the 

appliance connector/power cord when 

a chassis used by itself, but when 

installed into a rack or system the dis-

connect may be impaired and must be 

considered part of the installation.

Do Not Modify the Instrument

Do not install substitute parts or per-

form any unauthorized modification to 

the product. Return the product to an 

Keysight Sales and Service Office to 

ensure that safety features are main-

tained.

In Case of Damage

Instruments that appear damaged or 

defective should be made inoperative 

and secured against unintended oper-

ation until they can be repaired by 

qualified service personnel

Do NOT block vents and fan exhaust: 

To ensure adequate cooling and venti-

lation, leave a gap of at least 50mm 

(2") around vent holes on both sides of 

the chassis. 
Do NOT operate with empty slots: To 

ensure proper cooling and avoid dam-

aging equipment, fill each empty slot 

with an AXIe filler panel module.
Do NOT stack free-standing chassis: 

Stacked chassis should be rack-

mounted. 
All modules are grounded through the 

chassis: During installation, tighten 

each module's retaining screws to 

secure the module to the chassis and 

to make the ground connection.

Operator is responsible to maintain 

safe operating conditions. To ensure 

safe operating conditions, modules 

should not be operated beyond the full 

temperature range specified in the 

Environmental and physical specifica-

tion. Exceeding safe operating condi-

tions can result in shorter lifespan, 

improper module performance and 

user safety issues. When the modules 

are in use and operation within the 

specified full temperature range is not 

maintained, module surface tempera-

tures may exceed safe handling condi-

tions which can cause discomfort or 

burns if touched. In the event of a 

module exceeding the full temperature 

range, always allow the module to cool 

before touching or removing modules 

from the chassis.

 

iv

 

Содержание M9187A

Страница 1: ...Keysight M9187A PXI Digital IO Module Startup Guide...

Страница 2: ......

Страница 3: ...NOTICE IN FUTURE EDI TIONS FURTHER TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW KEYSIGHT DISCLAIMS ALL WAR RANTIES EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAIN...

Страница 4: ...erform proce dures involving cover or shield removal unless you are qualified to do so DO NOT operate damaged equipment Whenever it is possible that the safety protection features built into this prod...

Страница 5: ...k regulations under the terms of the Radio Commu nication Act of 1992 ICES NMB 001 indicates that this ISM device complies with the Canadian ICES 001 This symbol represents the time period during whic...

Страница 6: ...vi...

Страница 7: ...ep 4 Connect the Controller to the PXI Chassis 16 Using a remote PC 16 Using an embedded computer 16 Step 5 Install the Module in the PXI Chassis 17 Step 6 Verify Operation of the Module 19 Run Keysig...

Страница 8: ...viii Keysight M9187A PXI Digital IO Module Startup Guide...

Страница 9: ...puts can be connected directly to the outputs without risk of damage allowing module configuration for operation as 32 channels of independent input and output or as 32 channels of configurable I O Us...

Страница 10: ...t and heel strap combination Both types when used together provide a significant level of ESD protection Of the two only the table mat and wrist strap combination provides adequate ESD protection when...

Страница 11: ...Support information on the product web page at www keysight com find pxi dio 3 Write the following information on a tag and attach it to the malfunctioning equipment Name and address of owner A Post O...

Страница 12: ...Keysight IO Libraries Suite version 16 0 or later Any other accessories that you ordered cables connectors etc Step 3 Install the Software on the Host Computer The following illustration shows typical...

Страница 13: ...ncludes 1GB available for Microsoft NET Framework 3 5 SP1 100MB for Keysight IO Libraries Suite 1 5 GB available hard disk space includes 1GB available for Microsoft NET Framework 3 5 SP12 100MB for K...

Страница 14: ...Expert the IVI Shared Components and the VISA Shared Components 1 Insert the Automation Ready CD with Keysight IO Libraries Suite into the CD ROM drive of your system controller Wait a few seconds fo...

Страница 15: ...Microsoft development environments e g Visual Studio C C C Visual Basic Keysight VEE MATLAB or National Instruments Labview Accept the license terms and click Next Keysight Software License Agreement...

Страница 16: ...top With a Keysight Y1200A cable connect the adapter to the System Interface module on the PXI chassis Using an embedded computer Refer to the figure above If you are using an embedded controller in t...

Страница 17: ...lacing the module card edges into the front module guides top and bottom Slide the module to the rear of the chassis and assure that the injector ejector handle is pushed down in the unlatched downwar...

Страница 18: ...or laptop PC power on the computer Choose the default option for any Found New Hardware dialogs Chassis Power Down Process If you are using a remote controller linked to the M9021A Cable Interface yo...

Страница 19: ...xpert Locate your interfaces and instruments in the Keysight Connection Expert Explorer Pane The following graphic shows the Connection Expert screen Select the module in the center pane Instrument I...

Страница 20: ...Allow Control check box upper left corner before you can actually set the digital output channels If Allow Control is not checked the Output Channels section of the SFP display is grayed out Selectin...

Страница 21: ...to the single slot 3U form factor 100mm by 160mm 3 94 in by 6 3 in Eurocard standard These modules meet PXI Specification 2 2 Local bus trigger bus and star trigger are not included The modules also...

Страница 22: ...h Side driver device The outputs can also be tri stated off For serial input the module s 32 input channels are attenuated then multiplexed through an analog multiplexer A programmable threshold detec...

Страница 23: ...4 shows the M9187A front panel connector as viewed from the module front panel and connector pinout Table 1 on page 25 lists the front panel pin out connections by signal channel number Figure 3 shows...

Страница 24: ...utput 20 45 Output 19 26 Output 18 6 Output 17 66 Output 24 46 Output 23 27 Output 22 7 Output 21 67 Output 28 47 Output 27 28 Output 26 8 Output 25 68 Output 32 48 Output 31 29 Output 30 9 Output 29...

Страница 25: ...37 Input 27 56 Input 28 76 Input 29 18 Input 30 38 Input 31 57 Input 32 77 Output 1 2 Output 2 22 Output 3 41 Output 4 61 Output 5 3 Output 6 23 Output 7 42 Output 8 62 Output 9 4 Output 10 24 Output...

Страница 26: ...for the screw terminals is 0 12 0 15 Nm The following diagram shows the pinout of the connector block The pin numbers match the pin numbers on the module s front panel connector Refer to Disassemblin...

Страница 27: ...pper tin 26AWG with a 3A current rating Nominal resistance is 0 2 M The Y1187A is 1m long and the Y1188A is 2m long Figure 5 Y1187A Y1188A 78 Pin Male to Female Connector Cable Voltages greater than 6...

Страница 28: ...the strain relief bar Refer to Figure 7 4 Insert your cable under the strain relief bar and retighten the two strain relief screws 5 To reassemble the clamshell make certain the screw locks are in pl...

Страница 29: ...non condensing humidity The maximum humidity is 95 at 40 C Do not use in locations where conductive dust or electrolytic salt dust may be present These modules should be operated in an indoor environ...

Страница 30: ...30 Keysight M9187A PXI Digital IO Startup Guide Electrical Operating Conditions...

Страница 31: ...tems Keysight is a founding member of the LXI consortium www pxisa org PCI eXtensions for Instrumentation PXI modular instrumentation delivers a rugged PC based high performance measurement and automa...

Страница 32: ...This information is subject to change without notice Keysight Technologies 2014 Edition 3 December 2014 M9187 90001 M9187 90001 www keysight com...