Separate a Connection
To avoid lateral (bending) force on the connector mating plane surface, always
support the devices and connections.
1
Use an open-end wrench to prevent the device body from turning.
2
Use another open-end wrench to loosen the connector nut.
3
Complete the separation by hand, turning only the connector nut.
4
Pull the connectors straight apart without twisting, rocking, or bending either
of the connectors.
CAUTION
Turn the connector nut, not the device body. Major damage to the center
conductor can occur if the device is twisted.