1
Getting Started
22
Keysight M9164x and M9165x Startup and Service Guide
– Slide the module to the rear of the chassis and ensure that the injector/
ejector handle is pushed down in the unlatched (downward) position.
– Slide the module completely into the chassis. When you begin to feel
resistance, push up on the injector/ejector handle to fully inject the
M9164x/M9165x into the chassis.
5
Latch the module by pulling up on the injector/ejector handle, and secure the
front panel to the chassis using the module front-panel mounting screws.
6
Tighten the screws on the module front panel. Performance may suffer if the
screws are not tightened properly.
7
Install the System Interface Card in the chassis. (Refer to the System Interface
Card documentation for more information about the installation).
8
Verify that the PXI chassis fans are operable and free of dust and other
contaminants that may restrict air flow.
9
Install all chassis covers and filler panels after installing the M9164x/M9165x.
Missing filler panels may disrupt necessary air circulation in the chassis.
10
Connect the System Interface Card in the chassis to the host PC.
11
Plug in and power up the PXI chassis.
12
Power up the host PC.
NOTE
Warm up time of 5 minutes is recommended, after cold turn-on and after power
cycle.
Содержание M9164A
Страница 1: ...Keysight M9164A B C and M9165A B C PXIe Solid State Switch Module Startup and Service Guide...
Страница 10: ...10 Keysight M9164x and M9165x Startup and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 12: ...12 Keysight M9164x and M9165x Startup and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 14: ...14 Keysight M9164x and M9165x Startup and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...