Keysight M809256PB OIF CEI-56G Rx Test Automation Application User Guide
59
OIF CEI-56G VSR PAM4 Calibrations
4
SJ Calibration
Overview
The SJ Calibration is performed to calibrate the Sinusoidal Jitter.
Connection Diagram
Connect the instruments as shown in
for both VSR standards.
Figure 34
SJ Calibration connections for VSR Host & VSR Module
ID
Parameters
Refer to “
Calibration Parameters in Debug Mode
Procedure
This calibration uses the Clock/8 pattern.
The Victim Generator PJ1 is set until the desired SJ value is measured.
Results
This calibration returns the results for the set PJ1 versus measured SJ, in
graphical and tabular format.
Standard Name
Test ID
VSR Host
71104
VSR Module
271104