82
Keysight M809256PB OIF CEI-56G Rx Test Automation Application User Guide
5
OIF CEI-56G VSR PAM4 Tests
Figure 58
Multi Lane Stressed Input Test connections for VSR Module
ID
Dependencies
This test depends on the values configured for the following settings:
• All calibrations performed previously:
•
Crosstalk Calibration
•
UUGJ Calibration
•
SJ Calibration
Parameters
You may modify the following parameters for this test:
• Baud Rate
• Victim Analyzer Module
• Victim Analyzer Clock Source (select the source being used when
Victim Analyzer Module is a BERT Analyzer)
• Target Error Ratio
Standard Name
Test ID
VSR Host Multi-Lane
75101
VSR Module Multi-Lane
275101
NOTE
As mentioned earlier, UBHPJ Calibration is not required, since the SJ
Calibration sufficiently compensates for the UBHPJ Calibration.