Troubleshooting
Step 1. Shut down the E5080A.
Step 2. Disconnect all of the USB device from the USB ports.
Step 3. Connect the keyboard to the E5080A.
Step 4. Press the standby switch of the E5080A to turn it ON.
Step 5. When the screen as shown in the figure below appears,
select
Instrument Image Recovery System
and press
[Enter]
.
Figure 3
Instrument Image Recovery System selection screen
N O T E
After several seconds, the next screen appears automatically even
if you do not press any key.