Installation
1
Network Analyzer
13
Environmental Requirements
Set up the E5080A where the following environmental
requirements are met.
Operating Environments
Ensure that the operating environment meets the following
requirements.
Ventilation Requirement
To ensure safety requirements, the specifications and
measurement accuracy of the analyzer, you must keep the
environmental temperature within the specified range by
providing appropriate cooling clearance around the analyzer or,
for the rack mount type, by forcefully air-cooling inside the rack
housing. For more information on environmental temperature to
satisfy the specifications and measurement accuracy of the
analyzer, see the specification in the
E5080A Datasheet
.
When the environmental temperature around the analyzer is kept
within the temperature range of the operating environment
specification (See Section “Operating Environments” on page 13),
the analyzer conforms to the requirements of the safety standard.
Furthermore, under that temperature requirement, the analyzer
still conforms to the requirements of the safety standard even
when placing the analyzer with cooling clearance as follows:
Temperature
0 °C to 40 °C
Temperature range at
the error-correction
23 °C ± 3 °C (< 1 °C deviation from the temperature
when performing the error-correction)
Humidity
20% to 80% at wet bulb temperature < +29 °C
(non-condensation)
Altitude
0 to 2,000 m (0 to 6,561 feet)
Vibration
0.21 G maximum, 5 Hz to 500 Hz
N O T E
Above environmental requirements are NOT for the specifications
and measurement accuracy of the analyzer, but for the operating
environment of the analyzer.