Installation and Verification
4
Keysight 87406Q Operating and Service Manual
31
Operating and Service Instruction
Operator’s check
The operator’s check is supplied to allow the operator to make a
quick check on the switch prior to use or if a failure is suspected.
Description
The coaxial matrix switch is connected to a network analyzer
configured for the S- parameter measurement. The network
analyzer may be set to sweep over the whole or selected frequency
range of the switch to be verified. The S- parameter measurement
is the best way to determine if the switch is working properly.
C A U T I O N
ESD exceeding the level specified in
“Environmental specifications”
or RF
power applied is greater than the maximum specified as in
“General
Specifications”
may cause permanent damage to the device.
Figure 4-1
Connection to perform quick check
NETWORK ANALYZER
C
Keysight Coaxial
Transfer Switch
Port 1
Port 2
POWER SUPPLY
Содержание 87406Q
Страница 1: ...Keysight 87406Q Low PIM Coaxial Matrix Switch Operating and Service Manual ...
Страница 2: ......
Страница 3: ...Keysight 87406Q Operating and Service Manual 1 ...
Страница 12: ...10 Keysight 87406Q Operating and Service Manual 1 Introduction This page is intentionally left blank ...
Страница 30: ...28 Keysight 87406Q Operating and Service Manual 3 Specifications This page is intentionally left blank ...