24
Keysight 87406Q Operating and Service Manual
3
Specifications
Environmental specifications
The low PIM switches are designed to fully comply with Keysight’s
product operating environmental specifications.
Parameter
Specification
Temperature
•
Operating
•
Storage
•
Cycling
–25
°
C to 75
°
C
–55
°
C to 85
°
C
–50
°
C to 150
°
C, 10 cycles
Humidity
•
Operating
•
Storage
•
Condensation
40
°
C/95% RH, 5 days
65
°
C/90% RH, 24 hours
40
°
C/95% RH
Shock
•
Non-operating:
•
Half-sine
•
Transportation
•
Operating
500 G at 0.5 ms, 3 drops/direction
50 G Vibration: 8 m/s
±
10%
50 G at 6 ms, 6 directions
Vibration
•
Operating
•
Survival
•
Random
7 G rms, 5 to 2000 Hz at 0.25 in p-p
20 G rms, 20 to 2000 Hz at 0.06 in p-p, 4 min/cycle, 4 cycles/axis
7 G rms, 50 to 2000 Hz, 15 min/axis
ESD immunity
•
Direct discharge
•
Air discharge
6 kV (to outer conductor)
15 kV (to outer conductor)
RFI
Radiated emission per CISPR 11
Magnetic field
•
Operating emission
•
Operating immunity
AC magnetic emission (1.88 G rms)
DC magnetic emission (5 G)
30 A/M rms at 47 Hz, 50 Hz, 60 Hz, and 189 Hz
150 A/M rms at 47 Hz and 189 Hz