Operating and Service Guide 85059-90015 9
85059G 1.0 mm Connector Gage
Specifications
Table 3
Example when measuring adapters
Device
Typical Pin Depth
Measurement
Uncertainty
a
Observed Pin Depth
Limits
b
Guard Banded Pin Depth
Limits
c
Adapter
–0.0002 to –0.0010 in
–0.0040 to –0.0240 mm
±0.0002 in
±0.0040 mm
–0.0000 to –0.0012 in
–0.0000 to –0.0280 mm
–0.0004 to –0.0010 in
–0.0004 to –0.0008 mm
a. Calculated with coverage factor k
=
2.
b. Values observed within this range could be within the typical range due to measurement uncertainty.
c. Limits shown are guard banded by measurement uncertainty to ensure measured vaues are actually within the typical range.