Recommended Test Equipment
The following table is a list of the test equipment required to service this
instrument. The table indicates the critical specification of the test
equipment and for which procedure the equipment is necessary. Equipment
other than the recommended model may be used if it satisfies the critical
specification listed in the table.
Table 5
Recommended Test Equipment
Equipment
Required
Critical Specifications
Recommended
Model/Part
Use
DVM
Accuracy,
±
0.1%
3457A
P,A,T
Resistor
511
Ω
, 0.5 W, 1%
0757-0814
A
Resistor
50
Ω
, 10 W, 1%
0811-3707
A
P = Performance Tests, A = Adjustments, T = Troubleshooting
Service Strategy
The 1143A Probe Offset Control and Power Module consists of simple power
supplies and op-amp variable current sources. Circuitry is simple and
components are available off the shelf, so the service strategy is
component-level repair. The troubleshooting, parts lists, theory of operation,
and schematics support this repair level.
To clean the instrument
Use mild soap and water to clean the instrument. Harsh soaps will damage
the water-based paint finish of the instrument.
Service
General Information
23
Содержание 1143A
Страница 8: ...8 ...
Страница 16: ...16 ...
Страница 18: ...18 ...
Страница 43: ...Figure 6 Power and Control Assembly Component Locator Service Troubleshooting and Repair 43 ...
Страница 46: ...46 ...
Страница 48: ...Figure 8 Schematic 1 Line Input and Positive Supply 1 Service Troubleshooting and Repair 48 ...
Страница 49: ...Figure 9 Schematic 2 Negative Supply 2 Service Troubleshooting and Repair 49 ...
Страница 52: ...52 ...