To prepare the equipment
1
Turn off the front-panel power switch and remove the power cord.
W A R N I N G
When power is applied there are hazardous voltages inside the instrument.
Observe reasonable safety precautions in order to avoid injury or death.
2
Disconnect any probes from the front panel.
3
With a T10 Torx driver, remove 4 flathead screws and remove the top
chassis.
4
Check that the rear-panel line select switch is set for the correct line
voltage.
5
Connect the input power and turn on the front panel power switch.
To adjust power supply voltages
This procedure adjusts the power supply voltages. The adjustment adjusts
the
+
17.3 V supply directly. The
−
17.3 V supply tracks the
+
17.3 V supply.
Equipment Required
Equipment
Critical Specification
Recommended
Model/Part
Digital Multimeter
Better than 0.1% accuracy
3458A
Make sure there is no load on the supplies.
Use the adjustment locator on the previous page to find the test points and
adjustments.
1
Connect the DVM between TP5 (ground) and TP1 (
+
17.3V).
2
Adjust R38 (Positive Voltage Adjust) for a reading of 17.3V
±
50 mV.
3
Connect the DVM between TP5 (ground) and TP2 (
−
17.3V).
4
Check for a reading of
−
17.3V
±
150 mV.
If the supplies cannot be adjusted into tolerance, go to the "Troubleshooting and
Repair" section of this chapter.
Service
Making Adjustments
29
Содержание 1143A
Страница 8: ...8 ...
Страница 16: ...16 ...
Страница 18: ...18 ...
Страница 43: ...Figure 6 Power and Control Assembly Component Locator Service Troubleshooting and Repair 43 ...
Страница 46: ...46 ...
Страница 48: ...Figure 8 Schematic 1 Line Input and Positive Supply 1 Service Troubleshooting and Repair 48 ...
Страница 49: ...Figure 9 Schematic 2 Negative Supply 2 Service Troubleshooting and Repair 49 ...
Страница 52: ...52 ...