Test Steps and Description
Connect Request for
Bluetooth
®
5
This step must be a child step of
. This step acquires
Bluetooth
®
5 Signaling
PER of a connect request device.
Basic and Advanced Settings:
– View the descriptions for PER Limit, Packet Count, Repeat, and Early Fail Count in
– Response Iteration: Set the number of responses (1 to 10) to check for each connected PER method.
– Reconnect Retries: Set the number of times to reestablish connection to the DUT.
– Reconnect Power: Display the output power to reestablish connection.
– PER Method: Select one method out of the 19 options listed in the drop-down list. The default
settings are set as the Auto Detect method and three response iterations. When the default settings
are used, the radio will scan the devices based on the first connection method listed (UUID handle)
until it finds a method that reports 'Pass' consecutively for three response iterations.
– Check Supported PER Methods: Select the check box to allow the 34999B module to determine the
PER methods your DUT supports or manually select the Active Scan or Connect Request options.
– PHY: Choose one or more variants of Physical Layer (PHY) such as LE1M, LE2M, CodedS2 and
CodedS8. Generally,
Bluetooth
®
4.2 devices only support LE1M while certain
Bluetooth
®
5 devices
can support all four options. For the BLE 4.2 Signaling measurement suite, only the LE1M PHY is
available.
When a PER reading exceeds the limit, the test step to report ‘Fail’ verdict regardless of the number of
repeats. This test step will also report the Scan Response Bytes and Scan Data Bytes described below.
– Scan Response Bytes: Total bytes of device's scan data.
– Scan Data Bytes: Content of device's scan data.