U3042AM04
Controlling the Test Set with N5222
/32/42A/B
Controlling the Test Set with N5222
/32/42A/B
This section will describe how to operate the Test Set with the analyzer.
The internal firmware of the analyzer has not been modified for this Test Set option.
Power levels may differ from those indicated on the network analyzer when the Test
Set is connected.
The Test Set is considered a “slave” instrument. A PNA-X or PNA-L must be used to
control the Test Set. There are three methods to control the Test Set. Multiport mode
is recommended due to calibration and ease of use.
•
“Multiport Mode (Option 551)” on page
•
•
Typeface Key Conventions
The following key conventions are used throughout this document.
•
[HARDKEYS] are labeled front panel keys
•
SOFTKEYS are indicated on the instrument display
•
(Italicized in parenthesis)
are menu paths for the 'B' model analyzer
Sweep Setup for Multiport and Standalone Modes
When the Test Set is connected to the network analyzer, it is recommended that the
analyzer’s Sweep Setup be configured to Stepped Sweep before calibrating. This is
slower than the Analog Sweep, but is more accurate due to the extra electrical length of
the Test Set and test port cables. Stepped Sweep is available on all models.
1.
On the analyzer select STIMUL
U
S > Sweep > Sweep Setup.
2.
Select Stepped Sweep-sweep moves in discrete steps.
3.
Set the Dwell Time to 5
m
s > OK.
Figure
2
3
Sweep Setup
25
Keysight U3042AM04 User's and Service Guide