U3042AM04
Controlling the Test Set with
N5222/32/42A/B
Multiport Mode (Option 551)
Multiport mode selects the test set file that will enable the analyzer to control the
test set. Multiport mode allows you to complete a N-Port calibration using the Cal
Wizard application in the analyzer. Refer to the Help system for more information.
The U3042AM04 Option 129 (External Source Path Switch) will not operate in
multiport mode (Option 551).
How to Access Multiport Mode
1.
The Option 551 must be installed for Multiport capability. To access the multiport
application select [Utility] > System > Configure > Multiport Capability
-OR-
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...)
.
2.
Select U3042AM04 (4-Port System) from the test set drop-down menu and select
Restart as a multiport PNA with this testset > OK. The analyzer will restart the
network application with the test set interface features.
Figure 2
4
If the U3042AM04 is not available in the drop-down list, it will be necessary for you
to copy the required test set file to the analyzer's hard drive. The current version of
the test set files are available on the web at
http://na.support.keysight.com/
multiport
. Copy the appropriate file to
c:\program files\Keysight\Network Analyzer
\testsets
directory.
Selecting Multiport Mode
Keysight U3042AM04 User's and Service Guide
26