U3042AM04
Cal Kit Operational Check
1-Port Calibration and Verification Procedure
1.
Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing.
Torque to 8 in-lb. For further information refer to the Help menu.
2.
Perform a 1-Port Calibration on Port 1. On the analyzer, select
Response > Cal >
Start Cal > Calibration Wizard
-OR-
(Response > Cal > Other Cals)
.
a.
If using a mechanical cal kit, select SmartCal (Guided Calibration) > Next.
b.
If using an ECal module, connect the ECal to a PNA USB port. Select
Use
Electronic Calibration (ECal) > Next.
3.
Continue following the Cal Wizard prompts. In the Select Calibration Ports and
ECal Module dialog box click Clear All, then select Port 1 > Next > Measure.
4.
Ensure the Cal Kit you are using is indicated on the right side of the window.
Figure 4
2
1-Port Calibration
5. Continue to follow the prompts until the "Calibration Completed" dialog box
appears.
6. At the Calibration Completed prompt, select Save As User Calset, type the name
999.1. Overwrite the Calset if it already exists and Save.
If you do not have a keyboard, select
Save As User Calset
>
Edit Name
and save as
999.x
.
X is the port number you are calibrating. Use the numeric keypad on the
analyzer front panel to enter “999.1.”
Figure 4
3
Calibration Complete