U3042AM04
General
Specifications
General Specifications
Specifications for the Test Set are characteristic for the system performance of the
analyzer and Test Set. Actual performance of the system is based on the customers
analyzer that is used with the Test Set. A functional certificate is only offered for the
Test Set.
When connected to a analyzer, the Test Set will degrade the performance at the test
ports. The internal switch paths reduce test port power to the receivers. This affects
the test port power of the analyzer and also reduces dynamic range. The reflection
tracking values measured in the
“Cal Kit Operational Check” on page
can be
subtracted from the analyzers dynamic range to determine the approximate
performance of the system.
• 87104-60045 (switch, x8)
Power Requirements
Verify that the required ac power is available before installing the Test Set to the
analyzer.
•
100/120/220/240 V (50/60Hz)
•
The instruments can operate with mains supply voltage fluctuations up to ± 10%
of the nominal voltage.
•
Air conditioning equipment (or other motor–operated equipment) should not be
placed on the same AC line that powers the Test Set and analyzer.
•
U3042AM04 maximum power is 350 W.
This is a Safety Class I Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall be only be inserted in a
socket outlet provided with a protective earth contact. Any interruption of the
protective conductor inside or outside of the instrument, is likely to make the
product dangerous. Intentional interruption is prohibited.
The life expectancy of the following switches are 5 million cycles. When making
measurements using more than one port, single sweeps are recommended to minimize
switch cycles.
Definitions for Specifications
Specifications describe the warranted performance of calibrated instruments that
have been stored for a minimum of 2 hours within the operating temperature range of
0 to 55 °C, unless otherwise stated, and after a 45 minute warm-up period. Data
represented in this document are specifications unless otherwise noted.
Characteristics describe product performance that is useful in the application of the
product, but that is not covered by the product warranty. Characteristics are often
referred to as Typical or Nominal values.
•
Nominal describes representative performance that is useful in the application of
the product when operated over a 20 to 30 °C temperature range. Nominal
performance is not warranted.
•
Typical describes characteristic performance, which 80% of instruments are
expected to meet when operated over a 20 to 30 °C temperature range. Typical
performance is not warranted.