Keysight U3020AS22 User’s and Service Guide
2
0
U3020AS22
Controlling the Test Set
LAN Configuration Commands
IP Address Command
This command assigns a static Internet Protocol (IP) address for the U3020A.
Contact your network administrator for the valid IP address to use for your
instrument.
If you change the IP address, you must cycle power on the U3020A to activate
the new address.
Syntax
SYSTem:COMMunicate:LAN:IPADdress <address>
SYSTem:COMMunicate:LAN:IPADdress?
You can also queries the U3020A for the IP address it was assigned to.
Example
The following command sets the IP address:
SYST:COMM:IPAD 169.254.149.35
The following query returns the IP address currently being used by the
instrument (quotes are also returned).
SYST:COMM:LAN:IPAD?
Typical Response: "169.254.149.35"
Auto IP Address Command
This command disable or enable the use of Auto-IP standard to automatically
assign an IP address to the U3020A when on a network that does not have
DHCP servers.
Syntax
SYSTem:COMMunicate:LAN:AUTOip {OFF|0|ON|1}
SYSTem:COMMunicate:LAN:AUTOip?
Example
The following command disable the Auto-IP:
SYST:COMM:LAN:AUTOIP OFF
The following query returns the current Auto-IP setting:
SYST:COMM:LAN:AUTOIP?
Typical Response: 0