Operating the Power System 4
Keysight N6700 User’s Guide
93
102.4 microseconds:
204.8 microseconds:
409.6 microseconds:
409.6 microseconds:
1 parameter (voltage or current)
2 parameters (current)
4 parameters (min+max+current)
with 20 µs time interval resolution)
Up to 8 parameters with 40 µs time interval
resolution
You can change the time interval resolution as follows:
Front Panel:
SCPI Command:
Not Available
To set the resolution to 20 microseconds:
SENS:SWE:TINT:RES RES20,(@1)
To set the resolution to 40 microseconds:
SENS:SWE:TINT:RES RES40,(@1)
If the specified integration period is at or near the minimum logging
intervals, the data format must be specified as binary. If binary format is
not specified, the data will be in ASCII format and the minimum logging
intervals will typically be up to five times longer than what can be
achieved with binary format.
Front Panel:
SCPI Command:
Not Available
To set the data format to binary:
FORM[:DATA] REAL
Select the Data Logger Trigger Source
The TRIGger:ELOG command generates an immediate trigger regardless
of the trigger source. Unless you are using this command, select a
trigger source from the following:
BUS
Selects GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
EXTernal
Selects any pin configured as a trigger input as the trigger source.
IMMediate
Selects the immediate trigger source. This
triggers the data logger
immediately when initiated.
PIN<n>
Selects a pin on the digital port. <n> specifies the pin number. The pin
must be configured as a Trigger Input to be used as a trigger source
(see Appendix C).
Use the following commands to select one of the available trigger
sources:
Front Panel:
SCPI Command:
Not Available
TRIG:ELOG:SOUR BUS,(@1)
or
TRIG:ELOG:SOUR IMM,(@1)
or
TRIG:ELOG:SOUR EXT,(@1)
or
TRIG:ELOG:SOUR PIN<n>,(@1)