4 Operating the Power System
68
Keysight N6700 User’s Guide
Synchronizing Output Steps
The transient system lets you step the output voltage and current up or
down in response to triggered events. To generate a triggered output
step you must:
1.
Enable the output to respond to trigger commands.
2.
Set the voltage or current trigger levels.
3. Select the transient trigger source.
4. Initiate the trigger system and provide a trigger signal.
Enable the Output to Respond to Trigger Commands
First, you must enable the output to respond to trigger commands.
Unless an output is enabled to respond to triggers, nothing will happen
even if you have programmed a trigger level and generated a trigger for
the output.
Use the following commands to enable an output to respond to triggers:
Front Panel:
SCPI Command:
Select Transient\Mode.
For voltage step triggering, set the
Voltage mode to Step. For current
step triggering, set the Current
mode to Step. Then press Select.
To enable the voltage function
on output 1 to respond to
triggers, use:
VOLT:MODE STEP,(@1)
To enable the current function
on output 1 to respond to
triggers, use:
CURR:MODE STEP,(@1)
NOTE
In Step mode, the triggered value becomes the immediate value when
the trigger is received. In Fixed mode, trigger signals are ignored; the
immediate values remain in effect when a trigger is received.
Set the Voltage or Current Trigger Levels
Next, use the following commands to program an output trigger level.
The output will go to this level when the trigger is received.
If you have a model that has multiple ranges, the selected triggered
voltage and current settings must be within the same range that the
output channel is presently operating in.
Front Panel:
SCPI Command:
Select Transient\Step.
Select the Trig Voltage box to
set the voltage. Select the Trig
Current box to set the current.
Enter a value and press Select.
To set a voltage and current
trigger level for output 1 use:
VOLT:TRIG 15,(@1)
CURR:TRIG 1,(@1)