10
Functional Tests
Before Performing a Functional Test
Before Performing a Functional Test
1.
Ensure that you have the proper test equipment.
2.
Switch on the unit under test (UUT) and let it warm up (in accordance with
warm-up requirements in the instrument specifications).
3.
Allow sufficient warm-up time for the required test equipment (refer to
individual instrument documentation for warm-up specifications).
4.
Ensure that the analyzer’s frequency reference is set to Internal:
a.
Press the
Input/Output
, More,
Freq Ref In
keys.
b.
If the
Freq Ref In
softkey does
not
show Internal, press the
Freq Ref
In
softkey and select Internal.
5.
Following instrument warm-up, perform the auto align routine:
Press
System
,
Alignments
,
Align Now, All
.
Functional test accuracy depends on the precision of the test equipment
used. Ensure that all of the test equipment is calibrated before running a
functional test.