19
Keysight N9020A MXA Signal Analyzer
Functional Tests Guide
3
Frequency Readout Accuracy
Test Limits
Frequency Readout Accuracy is equivalent to the following equation:
±(0.25% × span + 5% × RBW + 2 Hz + 0.5 × horizontal resolution)
See results table for actual values.
The frequency readout accuracy is measured in several spans and center
frequencies that allow both internal analyzer synthesizer modes and prefilter
bandwidths to be tested. Frequency reference error is eliminated by using the
same frequency standard for the analyzer and signal source.
Item
Critical Specification
(for this test)
Recommended
Keysight Model
Adapter, Type-N(m), to
3.5 mm(f)
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.1:1
1250-1744
Adapter, 3.5 mm(f) to
3.5 mm(f)
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.1:1
83059B
Cable, 3.5 mm, 1 meter
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.4:1
11500E
Cable, BNC, 120 cm
Frequency: 10 MHz
10503A
Synthesized Sweeper
Frequency
: 10 MHz to 1.51 GHz
PSG