21
Frequency Readout Accuracy
Procedure
Procedure
1.
Configure the equipment as shown in
. Confirm the analyzer’s
built-in auto alignment has been performed within the past 24 hours.
2.
On the synthesized sweeper, press
PRESET
, then set the controls as
follows:
FREQUENCY
, 1505,
MHz
POWER LEVEL
, –10,
dBm
3.
Set up the signal analyzer by pressing:
Mode
,
Spectrum Analyzer
Mode Preset
Input/Output, More
,
Freq Ref In
,
External
FREQ Channel
,
Center Freq
, 1505,
MHz
SPAN X Scale, Span, 2990, MHz
Trace/Detector
,
More
,
Detector
,
Sample
Single
4.
Press
Restart
.
Press
Peak Search
on the analyzer. If the instrument is functioning
correctly, the marker reading in the active function block will be between
. Record the marker value in the Marker
Frequency Readout column in
5.
On the signal analyzer, change the span and center frequency as listed in
.
6.
Change the synthesized sweeper frequency to match the center frequency
of the analyzer.
7.
Repeat
through
until the Marker Frequency Readout column
of
Table 3-1
Frequency Readout Accuracy Results
Span
(MHz)
Center
Frequency
(MHz)
Minimum
Marker Frequency Readout
Maximum
2990
1505
1495.9 MHz
1514.1 MHz
127.2
1505
1504.56 MHz
1505.44 MHz
54.1
1505
1504.8122 MHz
1505.1878 MHz
7.95
1505
1504.97240 MHz
1505.0276 MHz
0.106
1505
1504.999630 MHz
1505.000370 MHz
1.98
517.59
517.58316 MHz
517.59684 MHz
1.98
832.50
832.49316 MHz
832.50684 MHz