27
Keysight N9020A MXA Signal Analyzer
Functional Tests Guide
5
Amplitude Accuracy at 50 MHz
Test Limits
Amplitude Accuracy should remain within 1.13 dB of the measured source
value across the range of source levels and changes in resolution bandwidth.
The Preamp (option P03, P08, P13, P26) should remain within
±
1.3 dB of
measured values.
A synthesized sweeper is used as the signal source for the test. The source
amplitude is varied using the signal source amplitude control and external step
attenuator. The resolution bandwidth is also varied on the signal analyzer. The
source amplitude is measured by the power meter and signal analyzer at each
setting, and the values compared. The difference between each pair of
measurements indicates the amplitude accuracy.
Item
Critical Specifications
Recommended
Keysight Model
Adapter
Type-N(m), to 3.5 mm(m)
1250-1743
Adapter
3.5 mm(f) to 3.5 mm(f)
83059B
Attenuator, 10 dB Step
Range: 10 to 30 dB
Frequency: 50 MHz
Accuracy
: ±0.5 dB
8495A
Attenuator, 30 dB
Accuracy
: < 0.5 dB at 50 MHz
11708A
Cable
3.5 mm, 1 meter
11500E
Cable
BNC, 120 cm
10503A
Power Meter
Compatible with power sensor
E4418B
Power Sensor
Amplitude Range
:
−
25 dBm to 10 dBm
8485A
Power Splitter
3.5 mm (f) connectors
6 dB loss
11667B
Synthesized Sweeper
Typical Temperature Stability
: 0.01 dBc/
°
C
PSG