14 Keysight M9383B Microwave Signal Generator Security Guide
Memory Declassification Procedure
M9383B VXG-m Microwave Vector Signal Generator
M9383B VXG-m Microwave Vector Signal Generator
Disk Drive Partitioning and Use
The drive is partitioned into 3 sections: C, D, and E.
— The C: partition contains the Windows 10 operating system and software
installed by Keysight. This is an Open System, this means you can install
additional software. Additional software should be installed on the C: drive,
however; only a limited set of software applications are tested for use with
the Keysight measurement software. If instrument repair is ever needed, the
Keysight version of the C: drive is the only part of the instrument software
restored by the Instrument Recovery process. You must reload any other
software that you have added in the instrument.
— The D: partition is reserved for data storage. The User Accounts that are
configured by Keysight have their Documents file system mapped to the D:
drive. This is for the convenience of backing-up the measurement data. You
should always backup the data on the D: drive to an external device. This
allows you to restore the data if you ever need to replace the disk drive.
— The E: partition is reserve for Keysight’s use. The primary use of the E: drive
is for housing the Calibration and Alignment data. Do not change or
overwrite the files on this drive. This could cause your instrument to not
meet specifications, or even to stop functioning correctly. Do not use this
drive for data storage.