Memory Declassification Procedure

Some test equipment users have a need to “declassify” or “sanitize” their 
instruments for security purposes. This involves following a procedure to clear 
all user data from the instrument’s memory. The result is a sanitized instrument 
that can be removed from a secure area without any chance of classified data 
being recovered from it. This document details the internal memory locations 
of the M9383B VXG-m. It describes instrument security features and the steps 
necessary to declassify the products through memory sanitization or removal. 
For additional information on a particular product, the Keysight Instrument 
Security Database may be accessed here: 

www.keysight.com/find/security

For general information, the Keysight Aerospace and Defense web page may 
be found at 

www.keysight.com/find/ad

.

What you will find in this section:

“Security Terms and Definitions” on page 12

“Product Memory Sanitization” on page 13

“M9383B VXG-m Microwave Vector Signal Generator” on page 14

“M9043A Chassis and Frequency Reference” on page 18

“M9347A PXIe Direct Digital Synthesizer” on page 19

“M9312B PXIe Source Output” on page 20

“M9314B PXIe Upconverter” on page 21

“M9323A PXIe Amplifier” on page 22

“M9343A PXIe Digital Vector Modulator” on page 23

Содержание M9383B VXG-m

Страница 1: ...Security Guide Keysight M9383B VXG m Microwave Vector Signal Generator...

Страница 2: ...use modify reproduce release perform display or disclose commercial computer software or commercial computer software documentation No additional government requirements beyond those set forth in the...

Страница 3: ...s and Service Offices 9 Memory Declassification Procedure Security Terms and Definitions 12 Product Memory Sanitization 13 M9383B VXG m Microwave Vector Signal Generator 14 Disk Drive Partitioning and...

Страница 4: ...4 Keysight M9383B Microwave Signal Generator Security Guide Contents...

Страница 5: ...sible for the use and maintenance of equipment for ensuring that the equipment is operated within its specifications and operating limits and for ensuring operators are adequately trained Operators us...

Страница 6: ...any unknown circuit before measuring Operators of this product must be protected from electric shock at all times The responsible body must ensure that operators are prevented access and or insulated...

Страница 7: ...re performing any maintenance disconnect the line cord and all test cables To maintain protection from electric shock and fire replacement components in mains circuits including the power transformer...

Страница 8: ...ed as Monitoring and Control instrumentation product Do not dispose in domestic household waste To return unwanted products contact your local Keysight office or for more information see http about ke...

Страница 9: ...Is your product software up to date Periodically Keysight releases software updates to fix known defects and incorporate product enhancements To search for software updates for your product go to the...

Страница 10: ...10 Keysight M9383B Microwave Signal Generator Security Guide Safety Information Contacting Keysight...

Страница 11: ...teps necessary to declassify the products through memory sanitization or removal For additional information on a particular product the Keysight Instrument Security Database may be accessed here www k...

Страница 12: ...ion procedures include memory sanitization or memory removal or both Keysight declassification procedures are designed to meet the requirements specified in DoD 5220 22 M Chapter 8 Sanitization As def...

Страница 13: ...sses for the following Keysight product models are covered in this document Instrument M9383B VXG m Microwave Vector Signal Generator PXI Modules M9043A PXI Chassis and Frequency Reference M9037A PXIe...

Страница 14: ...of the C drive is the only part of the instrument software restored by the Instrument Recovery process You must reload any other software that you have added in the instrument The D partition is rese...

Страница 15: ...documentation User Modifiable Yes Volatile No Memory Erase Processes To uninstall the PathWave instrument driver from the controller perform the relevant procedure below To erase 1 Select Start Contr...

Страница 16: ...ing information from the SSD SSD Removal and Installation Because it is virtually impossible to completely and selectively erase all user data on a hard drive without also destroying the operating sys...

Страница 17: ...oved from the M9037A do not attempt to power it up Always install the SSD before applying power to the M9037A If you do not then the SATA selection is eliminated from the boot option list If the SSD i...

Страница 18: ...date max assembly temperature and calibration data User Modifiable No Volatile No Memory Erase Processes None this is not user accessible Memory Type 2 Memory Type Flash Memory Memory Size 128 M Bit...

Страница 19: ...cal verify date max module temperature and calibration and alignment data User Modifiable No Volatile No Memory Erase Processes None this is not user accessible Memory Type 2 Memory Type Flash Memory...

Страница 20: ...perature and calibration and alignment data User Modifiable No Volatile No Memory Erase Processes None this is not user accessible Memory Type 2 Memory Type Flash Memory Memory Size 128 M Bit Memory F...

Страница 21: ...erature and calibration and alignment data User Modifiable No Volatile No Memory Erase Processes None this is not user accessible Memory Type 2 Memory Type Flash Memory Memory Size 128 M Bit Memory Fu...

Страница 22: ...dule temperature and calibration and alignment data User Modifiable No Volatile No Memory Erase Processes None this is not user accessible Memory Type 2 Memory Type Flash Memory Memory Size 128 M Bit...

Страница 23: ...verify date max module temperature and calibration and alignment data User Modifiable No Volatile No Memory Erase Processes None this is not user accessible Memory Type 2 Memory Type Flash Memory Mem...

Страница 24: ...This information is subject to change without notice Keysight Technologies 2020 Edition 1 June 2020 M9383 90021 www keysight com...

Отзывы: