76
4. Calibration / Deskew Procedure
Procedure
16
With the probe head correctly connected to the fixture, press OK in the
dialog box on the oscilloscope and wait for the calibration to finish. It
will report whether the calibration was successful or not.
17
Once the calibration has successfully completed, the DC Cal LED on
the probe amplifier will turn green indicating that the particular
combination of probe amplifier, probe head, and oscilloscope channel
input has been calibrated.
NOTE
Any time you use a new probe head type, a new probe amplifier, or a different
channel on the oscilloscope, the
DC Cal
LED will be amber, indicating that a DC
calibration is required.
18
Once the DC vertical calibration has successfully completed, select
the Calibrated Skew radio button from the Probe Calibration dialog
box on the oscilloscope. Then press the Start Skew Calibration...
button and follow the on-screen instructions for the skew calibration.
Содержание InfiniiMax III Series
Страница 1: ...Keysight InfiniiMax III Series Probes User s Guide...
Страница 4: ...4 Contents...
Страница 28: ...28 1 Using InfiniiMax III Series Probes Simplified InfiniiMax III Schematic...
Страница 64: ...64 2 InfiniiMax III Probe Heads N5444A InfiniiMax III 2 92 mm 3 5 mm SMA Probe Head...
Страница 100: ...100 5 Performance Verification Performance Test Record...
Страница 103: ...103 6 Performance Plots N2836A 26 GHz Solder In Head with N2803A N2836A 26 GHz Solder In Head with N2803A...
Страница 104: ...104 6 Performance Plots N2836A 26 GHz Solder In Head with N2803A...
Страница 105: ...105 6 Performance Plots N2836A 26 GHz Solder In Head with N2803A...
Страница 107: ...107 6 Performance Plots N2838A N5439A 25 GHz PCB ZIF Tip Head with N2803A...
Страница 108: ...108 6 Performance Plots N2838A N5439A 25 GHz PCB ZIF Tip Head with N2803A...
Страница 110: ...110 6 Performance Plots N5440A N5439A 28 GHz Ceramic ZIF Tip Head with N2803A...
Страница 111: ...111 6 Performance Plots N5440A N5439A 28 GHz Ceramic ZIF Tip Head with N2803A...
Страница 113: ...113 6 Performance Plots N5447A N5439A 28 GHz Ceramic High Sensitivity ZIF Tip Head with N2803A...
Страница 114: ...114 6 Performance Plots N5447A N5439A 28 GHz Ceramic High Sensitivity ZIF Tip Head with N2803A...
Страница 115: ...115 6 Performance Plots N5441A 16 GHz Solder In Head with N2803A N5441A 16 GHz Solder In Head with N2803A...
Страница 116: ...116 6 Performance Plots N5441A 16 GHz Solder In Head with N2803A...
Страница 117: ...117 6 Performance Plots N5441A 16 GHz Solder In Head with N2803A...
Страница 118: ...118 6 Performance Plots N5444A 28 GHz SMA 2 92 mm Head with N2803A N5444A 28 GHz SMA 2 92 mm Head with N2803A...
Страница 119: ...119 6 Performance Plots N5444A 28 GHz SMA 2 92 mm Head with N2803A...
Страница 121: ...121 6 Performance Plots N5445A 30 GHz Browser Head 1 mm span with N2803A...
Страница 122: ...122 6 Performance Plots N5445A 30 GHz Browser Head 1 mm span with N2803A...
Страница 128: ...128 T Tack putty 42 59 W wrist strap 10 Z ZIF Probe Head 38 connecting 44 ZIF tip 6 soldering 41...