148
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
11
Trigger Mode/Coupling
To select the Auto or Normal trigger mode
When the oscilloscope is running, the trigger mode tells the oscilloscope what to
do when triggers are not occurring.
In the
Auto
trigger mode (the default setting), if the specified trigger conditions are
not found, triggers are forced and acquisitions are made so that signal activity is
displayed on the oscilloscope.
In the
Normal
trigger mode, triggers and acquisitions only occur when the
specified trigger conditions are found.
To select the trigger mode:
1
Press the
[Mode/Coupling]
key.
2
In the Trigger Mode and Coupling Menu, press the
Mode
softkey; then select
either
Auto
or
Normal
.
See the following
“When to Use Auto Trigger Mode"
on page 149 descriptions.
You can also configure the
[Quick Action]
key to toggle between the Auto and
Normal trigger modes. See
“Configuring the [Quick Action] Key"
Triggering and the
Pre- and
Post-Trigger
Buffers
After the oscilloscope starts running (after pressing
[Run]
or
[Single]
or changing
the trigger condition), the oscilloscope first fills the pre-trigger buffer. Then, after
the pre-trigger buffer is filled, the oscilloscope starts searching for a trigger, and
sampled data continues to flow data through the pre-trigger buffer in a first-in
first-out (FIFO) manner.
When a trigger is found, the pre-trigger buffer contains the events that occurred
just before the trigger. Then, the oscilloscope fills the post-trigger buffer and
displays the acquisition memory. If the acquisition was initiated by
[Run/Stop]
, the
process repeats. If the acquisition was initiated by pressing
[Single]
, the acquisition
stops (and you can Pan and Zoom the waveform).
In either Auto or Normal trigger mode, a trigger may be missed if the event occurs
while the pre-trigger buffer is being filled. This may be more likely, for example,
when the horizontal scale knob is set to a slow time/div setting, such as
500 ms/div.
Trigger Indicator
The trigger indicator at the top right of the display shows whether triggers are
occurring.
In the
Auto
trigger mode, the trigger indicator can show:
Содержание DSO-X 2002A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Страница 344: ...344 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...
Страница 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...