Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
10
Triggers
•
Field1
and
Field2
— Trigger on the rising edge of the first serration pulse of
field 1 or field 2 (interlaced standards only).
•
All Fields
— Trigger on the rising edge of the first pulse in the vertical sync
interval.
•
All Lines
— Trigger on all horizontal sync pulses.
•
Line: Field1
and
Line:Field2
— Trigger on the selected line # in field 1 or field 2
(interlaced standards only).
•
Line: Alternate
— Alternately trigger on the selected line # in field 1 and field 2
(NTSC, PAL, PAL-M, and SECAM only).
9
If you select a line # mode, press the
Line #
softkey, then rotate the Entry knob
to select the line number on which you want to trigger.
The following table lists the line (or count) numbers per field for each video
standard.
Video Triggering
Examples
The following are exercises to familiarize you with video triggering. These exercises
use the NTSC video standard.
•
“To trigger on a specific line of video"
•
“To trigger on all sync pulses"
•
“To trigger on a specific field of the video signal"
•
“To trigger on all fields of the video signal"
•
“To trigger on odd or even fields"
To trigger on a specific line of video
Video triggering requires greater than 1/2 division of sync amplitude with any
analog channel as the trigger source. Turning the trigger
Level
knob in Video
trigger does not change the trigger level because the trigger level is automatically
set to the sync pulse tips.
Video standard
Field 1
Field 2
Alt Field
NTSC
1 to 263
1 to 262
1 to 262
PAL
1 to 313
314 to 625
1 to 312
PAL-M
1 to 263
264 to 525
1 to 262
SECAM
1 to 313
314 to 625
1 to 312
Содержание DSO-X 2002A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Страница 344: ...344 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...
Страница 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...