Triggers
10
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
137
6
In the Video Trigger Menu, press the
Standard
softkey to set the video standard.
The oscilloscope supports triggering on the following television (TV) and video
standards.
7
Press the
Auto Setup
softkey to automatically set up the oscilloscope for the
selected
Source
and
Standard
:
•
Source channel vertical scaling is set to 140 mV/div.
•
Source channel offset is set to 245 mV.
•
Source channel is turned on.
•
Trigger type is set to
Video
.
•
Video trigger mode is set to
All Lines
.
•
Display
Grid
type is set to
IRE
(when
Standard
is
NTSC
) or
mV
(see
•
Horizontal time/division is set to 10 µs/div for NTSC/PAL/SECAM standards.
•
Horizontal delay is set so that trigger is at first horizontal division from the
left.
You can also press
[Analyze]> Features
and then select
Video
to quickly access the
video triggering automatic set up and display options.
8
Press the
Mode
softkey to select the portion of the video signal that you would
like to trigger on.
The Video trigger modes available are:
Standard
Type
Sync Pulse
NTSC
Interlaced
Bi-level
PAL
Interlaced
Bi-level
PAL-M
Interlaced
Bi-level
SECAM
Interlaced
Bi-level
Содержание DSO-X 2002A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Страница 344: ...344 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...
Страница 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...