background image

9

c. Press           ,            , and then press                    to configure to use 4-wire 
 connection. 

 

d. Press Ch1             to turn on Channel 1 Output relay.
 
e. Press            to perform a single point measurement.

4. Compare two results

a.  Compare two results to see the effect of 4-wire connection. The result with  

  4-wire connection is 1 Ohm, while the one with 2-wire connection is 1.6 Ohm.  

  The difference, that is 0.6 Ohm, should be the residual lead resistance on the  

  measurement cables.

LAB 2: Make an Accurate Low Resistance Measurement (continued)

(3) Press OK

(2) Press 2-WIRE

(1) Press

High Force

Low Force

B2901/02/11/12A

R

DUT

R

Lead

R

Lead

V

Meas

I

Source

a) Result with 2-wire connection

b) Result with 4-wire connection

High Force

Low Force

Low Sense

High Sense

B2901/02/11/12A

V

Meas

I

Source

R

DUT

R

Lead

A

B

The result with 2-wire connection includes the residual lead resistance R

Lead

Содержание B2901A

Страница 1: ...Keysight Technologies Detailed Demonstration Guide Making Resistance Measurement Using SMU Keysight B2901A 02A 11A 12A Precision Source Measure Unit Demonstration Guide...

Страница 2: ...plication software that make it easy for you to begin making productive measure ments immediately This demonstration guide shows how easily basic resistance measurement can be made using the Keysight...

Страница 3: ...ed tweezers 4 Clip the other lead of 1 k Resistor with the red gold plated tweezers Demonstration 1 Change View mode to Single View a Press repeatedly until Single View for Channel 1 is shown in the d...

Страница 4: ...o stop making measurements periodically 3 Turn off the channel output a Press Ch1 to turn off Channel 1 Output relay LAB 1 Make a Resistance Measurement continued Measurement parameters are updated pe...

Страница 5: ...Resistor with black gold plated tweezers Why is 4 wire connection required For very low resistance measurements where the residual lead resistance is comparable to the DUT resistance the 2 wire measu...

Страница 6: ...wire connection you can see the status indicator on GUI as below although no indicator can be seen on being configured to use 2 wire connection Demonstration 1 Reset the instrument a Press and then pr...

Страница 7: ...nd select and then press to configure to use 4 wire connection c Press then press to set Channel 1 V I Source Function to I Source If can t be found on the Assist keys press to change the keys d Press...

Страница 8: ...asure condition and calculates the resistance value e Press and set Channel 1 Limit value to 1 V f Rotate to select Resistance measurement operation and press to edit it Then press to set Resistance m...

Страница 9: ...connection is 1 6 Ohm The difference that is 0 6 Ohm should be the residual lead resistance on the measurement cables LAB 2 Make an Accurate Low Resistance Measurement continued 3 Press OK 2 Press 2 W...

Страница 10: ...rely on accurate measurements www keysight com go quality Keysight Technologies Inc DEKRA Certified ISO 9001 2008 Quality Management System Keysight Channel Partners www keysight com find channelpartn...

Отзывы: