background image

The Keysight Technologies, Inc. 

B2901A/02A/11A/12A Precision 

Source/Measure Units are compact 

and cost-effective bench-top Source/

Measure Units (SMUs) with the 

capability to output and measure 

both voltage and current. The 

B2901A/02A/11A/12A enables you 

to make a wide range of current 

versus voltage (IV) measurements 

more accurately and quickly 

than ever before. In addition, the 

B2901A/02A/11A/12A comes with 

an intuitive graphical user interface 

(GUI) and free PC-based application 

software that make it easy for you to 

begin making productive measure-

ments immediately. 
This demonstration guide shows how 

easily basic resistance measurement 

can be made using the Keysight 

B2901A/02A/11A/12A.
The demonstration includes not only 

a simple resistance measurement 

using 2-wire connection, but also a 

low resistance measurement using 

4-wire connection.

Required Instrument and Accessories

Keysight 11059A Kelvin Probe Set and Resistors are equipped as a demo kit 

with a demo unit of the Keysight B2900A Series of SMU.

Keysight 11059A Kelvin Probe Set

1 Ω Resistor

1 kΩ Resistor

Keysight B2901A/02A/11A/12A

Precision Source/Measure Unit

Introduction

Содержание B2901A

Страница 1: ...Keysight Technologies Detailed Demonstration Guide Making Resistance Measurement Using SMU Keysight B2901A 02A 11A 12A Precision Source Measure Unit Demonstration Guide...

Страница 2: ...plication software that make it easy for you to begin making productive measure ments immediately This demonstration guide shows how easily basic resistance measurement can be made using the Keysight...

Страница 3: ...ed tweezers 4 Clip the other lead of 1 k Resistor with the red gold plated tweezers Demonstration 1 Change View mode to Single View a Press repeatedly until Single View for Channel 1 is shown in the d...

Страница 4: ...o stop making measurements periodically 3 Turn off the channel output a Press Ch1 to turn off Channel 1 Output relay LAB 1 Make a Resistance Measurement continued Measurement parameters are updated pe...

Страница 5: ...Resistor with black gold plated tweezers Why is 4 wire connection required For very low resistance measurements where the residual lead resistance is comparable to the DUT resistance the 2 wire measu...

Страница 6: ...wire connection you can see the status indicator on GUI as below although no indicator can be seen on being configured to use 2 wire connection Demonstration 1 Reset the instrument a Press and then pr...

Страница 7: ...nd select and then press to configure to use 4 wire connection c Press then press to set Channel 1 V I Source Function to I Source If can t be found on the Assist keys press to change the keys d Press...

Страница 8: ...asure condition and calculates the resistance value e Press and set Channel 1 Limit value to 1 V f Rotate to select Resistance measurement operation and press to edit it Then press to set Resistance m...

Страница 9: ...connection is 1 6 Ohm The difference that is 0 6 Ohm should be the residual lead resistance on the measurement cables LAB 2 Make an Accurate Low Resistance Measurement continued 3 Press OK 2 Press 2 W...

Страница 10: ...rely on accurate measurements www keysight com go quality Keysight Technologies Inc DEKRA Certified ISO 9001 2008 Quality Management System Keysight Channel Partners www keysight com find channelpartn...